Colin BP-S510 Patient Monitor

Condition: Refurbished

Model Number:Colin BP-S510 Patient Monitor

Description:
Product Description
Leading-edge patient monitor features patented Colin NIBP technology
The BP-S510 has a sophisticated design that provides simple, user-friendly operation. Its advanced functionality makes it perfect for a wide variety of uses, including post-anesthesia care, ambulatory surgery, endoscopy, emergency rooms and more!

* Measures ECG (3/5 lead), heart rate, Colin NIBP, Nellcor SpO2 and respiration. Available with temperature, two-channel IBP, EtCO2, and a printer.
* Compact: Conveniently compact and feature-packed, the BP-S510 is great for areas where space is at a premium.
* Lightweight: Weighing only 12.1 lbs, the BP-S510 is easy to carry from ward to ward. This lightweight device makes it easy to move to stepdown, PACU, RR, etc.
* Battery Operation for Portability: Available with rechargeable batteries, the BP-S510 can be used during transport within your facility.
* Wide Viewing Angle: The BP-S510’s large LCD screen features a wide viewing angle. Each parameter is color-coded in bright and vivid colors, making the information easy to read even from a distance. You can also select from multiple views to highlight just the information you need.

Technical Specifications
Non-Invasive Blood Pressure (NIBP) – Patented Colin Technology

* Measurement technology Oscillometric method
* Measurement method Linear deflation method
* Pressure display range 10 to 300mmHg
* Pressure display accuracy Less than ±3mmHg
* NIBP Measurement range
* Adult/pediatric mode

SYS 60 to 250mmHg
MAP 45 to 235mmHg
DIA 40 to 200mmHg
Pulse rate 40 to 200bpm

* Pulse rate 40 to 240bpm
* NIBP accuracy AAMI SP-10:2002+A1:2003
* Pulse rate accuracy ±2% or ±2 beats

ECG (Electrocardiogram)

* Leads 3/5 Lead
* Lead I, II, III, aVR, aVL, aVF, V (Chest Lead)
* Heart Rate
* Measurement Range 20 to 250bpm
* Accuracy ±3bpm or ±5% (whichever is greater)

Respiration (im RR)

* Measurement method Trans-thoracic impedance
* Leads RA to LA
* Range 0, 3 to 120breaths/min
* Accuracy ±3 breaths/min

SpO2 (Pulse oximeter)

* Measurement method 2 wave length pulse wave type
* Range 1 to 100% SpO2
* Accuracy Adult/pediatric (When using SpO2 Disposable Sensor MAX-A) 70 to 100% SpO2 ±2 digits, 1 to 69% SpO2 unspecified
* Pulse Rate Range 20 to 250bpm
* Pulse Rate Accuracy ±3bpm
* Display update Less than 30 sec

Temperature Measurement (Option)

* Measurement method Thermistor
* Probe Type Thermistor probe YSI 400 series (2 channels) (Probes not included)
* Range 15 to 45°C (59 to 113°F)
* Display Accuracy ±0.1°C (25 to 45°C) (77 to 113°F) ±0.2°C (15 to less than 25°C) (59 to less than 77°F)
* Probe Accuracy ±0.1°C (±0.18°F)

IBP (Option)

* Measurement Method: Pressure Transducer Method
* Pressure Measurement Range – -50 to +300mmHG
* Pulse Rate Measurement Range – 20 to 250bpm
* Display Accuracy – ±3mmHG
* Standard – IEC60601-2-34:2000

CO2 (Option)

* Measurement method Non-dispersive Infrared Spectroscopy
* Range 0-100mmHg : 0-13.3kpa : 0-10%CO2
* Accuracy ±2mmHg or ±4% (whichever is greater)
* Respiration (awRR) Range 0 to 120 breaths/min
* Respiration (awRR) Accuracy ±1 breath/min

Recorder (Option)

* Type Thermal
* Weight 150g (0.33lbs)
* Resolution 8 dot/mm
* Number of channels 1 to 2 channels
* Paper Type Thermal
* Paper Width 50mm (2in)
* Paper Speeds 25.0mm/sec and 50.0mm/sec (1in/sec and 2in/sec)

General

* Dimensions 360(W) x 215(D) x 280(H)mm (14.2 x 8.5 x 11in)
* Weight 5.5kg (12.1lbs) (excluding optional configurations and accessories)
* Display – Screen Size 26.4cm (10.4in) measured diagonally, Screen Type/Color Liquid Crystal Display (LCD) color, Resolution 800 x 600 pixels
* Number of traces 4 or 6 waveforms
* Safety Standard IEC60601-1:1988+A1:1991+A2:1995
* Degree of protection (enclosures) IP21
* Type of protection Type CF – Applied part

Power Requirement

* AC Power 100-240V~50/60Hz, 0.5-1.3A
* Battery Ni-MH, 12V/3.8Ah (1 included)

Environmental

* Operating Temperature 10 to 40°C (50 to 104°F)
* Storage Temperature -20 to 60°C (-4 to 140°F)
* Operating Humidity 15 to 90% RH, non-condensing

Accessories
Most models: Reusable adult cuff, adult 3.5m cuff hose, ECG cable,
ECG 3-lead wires, SpO2 Durasensor® DS-100A, SpO2 extension cable
DOC-10, operation manual, power cord, rechargeable battery (1).
Available with: CO2 sampling set, printer paper, ECG 5-lead wires,
temperature probes, rolling stand, and a full size range of reusable and
disposable BP cuffs

Our Policy:
This equipment is guaranteed to be as we represent it in the description or your purchase price will be refunded upon return in as-shipped condition for a full refund of sale price. Please note that we record all serial numbers of each item that we ship. Also please note that we do not refund shipping, handling and packaging charges. You must notify us of any discrepancies immediately upon receipt of equipment to make arrangements prior to any return. This guarantee applies to U.S. shipping addresses within the lower 48 states only (All sales or shipments to other locations will be as-is and are final

Request Quote For This Item

Your comments or requests

*